Table 6: Calibration of Calibrator-Mode Resistance Functions
(RTD-Thermistor)
Sensor
Step
Select
1
RTD-100
2
RTD-100
3
RTD-100
4
1,000Ω
5
1,000Ω
6
100,000Ω
7
100,000Ω
8
Thermistor
9
Thermistor
10
Thermistor
* exclusive of noise.
2. CALIBRATION VERIFICATION
Calibration is verified with the same instrument hookups as used for calibra-
tion (ie. Figures 12 and 13 for thermocouple functions, Figure 14 for resis-
tance functions).
While one hookup serves to verify all resistance functions (RTD, thermistor,
ohms), a full check-out of thermocouple functions requires a set of calibration
cables for each thermocouple type.
To simplify the verification of thermocouple calibration, software has been
enhanced to reactivate the copper-mode used during calibration. This feature
is available on products that use the following (and later) software revisions:
32
Current
Unit
Source
Output
Input
1562.0°F
1mA
32.0°F
1mA
–328.0°F
1mA
900.00Ω
1mA
1.00Ω
1mA
90,000Ω
10µA
100Ω
10µA
293.0°F
1mA
77.0°F
10µA
–40.0°F
10µA
Adjust
Verify Reading*
P38
√
√
√
√
P51
√
√
√
√
DMM
0.39026v
0.10000
±0.00006v
0.01849
±0.00012v
0.90000
±0.00020v
0.00100
±0.00020v
0.90000v
0.00100
±0.00020v
0.04700
±0.00020v
0.02252
±0.00020v
0.75790
±0.00500v