Replacing The Fuses; Accessories - Megger AVO300 Series User Manual

Table of Contents

Advertisement

Available languages

Available languages

REPLACING THE FUSES

WARNING: To avoid electric shock, disconnect the test leads from any
source of voltage before removing the fuse/battery cover.
1. Disconnect the test leads from the meter and any item under test.
2. Open the fuse door by loosening the screw on the door using a cross
head screwdriver.
3. Remove the old fuse from its holder by gently pulling it out.
4. Install the new fuse into the holder.
5. Always use a fuse of the proper size and value (0.5 A/1000 V fast blow for
the 400 mA range, 10 A/1000 V fast blow for the 10 A range).
6. Put the fuse door back in place. Insert the screw and tighten it securely.
WARNING: To avoid electric shock, do not operate your meter until the
fuse door is in place and fastened securely.

ACCESSORIES

Included accessories
Standard Red/Black lead set with test probes
Optional Accessories
Standard lead set with test probes and croc clips
Fused lead set with test probes and croc clips
6220-779
6220-789
11

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the AVO300 Series and is the answer not in the manual?

Questions and answers

This manual is also suitable for:

Avo310

Table of Contents