Spring Adapter For The Water Connection; Adjusting The Upper Basket - Miele Laboratory Glassware Washer G 7883 Operating And Installation Instructions

Laboratory glassware washer
Table of Contents

Advertisement

Loading the machine
Spring adapter for the water
connection
The spring adapter for the water
connection must be properly engaged
when a basket or the injector wagon is
inserted into the machine. It must be
4 - 5 mm higher than the water inlet in
the machine.
If this is not the case, the adapter
needs to be adjusted as follows:
^ Loosen the locking ring a.
^ Move the adapter up to the proper
height b.
^ Tighten the locking ring c.
24

Adjusting the upper basket

The upper basket can be adjusted
above and below the middle position,
by 2 cm.
Depending on the height selected and
which inserts are used, labware of
various heights can be arranged in the
baskets.
To adjust the upper basket:
^ Pull the upper basket out as far as it
will go, lift it up off the runners and
remove.
^ Use a wrench to unscrew the roller
bearings on either side of the basket,
then reposition as required.