9. Set the signal generator as follows:
Frequency
...............................................................................................................
1300 MHz
Amplitude
..............................................................................................................
+
0.5 dBm
NOTE
This amplitude provides 75 mV rms to the Channel C Input when using the
10
dB attenuator.
VERIFY: The Counter displays 1300 MHz
+
20 Hz.
10. Record the measurement on the Performance Test Record Card, line 56.
Failure: Perform Section V, Adjustments, paragraphs 5-21 and 5-22.