Universal counter
Hide thumbs Also See for 5334B:
Table of Contents

Advertisement

2. Set the HP 5334B as follows:
Reinitialize the 5334B.
FUNCTION
...........................................................................................................
FREQ C
CHANNEL C SENSITIVITY control
..............................................................
fully CW
3. Connect the signal generator output to the HP 5334B Input
C through a 10 dB attenuator.
NOTE
The 10 dB attenuator is used here for impedance matching.
VERIFY: The Counter displays 90.0 MHz
+
2
H z .
4.
Record the measurement on the Performance Test Record Card, line 53.
5. Set the signal generator as follows:
Frequency
...............................................................................................................
200 MHz
Amplitude
..............................................................................................................
-13.5 dBm
NOTE
This amplitude provides 15 mV rms to the Channel C lnput when using the
10 dB attenuator.
VERIFY: The Counter displays 200.0 MHz
+
3 Hz.
6.. Record the measurement on the Performance Test Record Card, line 54.
7.
Set the signal generator as follows:
Frequency
...............................................................................................................
1000 MHz
Amplitude
..............................................................................................................
+
0.5 dBm
NOTE
This amplitude will provide 75 mV rms to the Channel C lnput when using
the 10 dB attenuator.
VERIFY: The Counter displays 1000 MHz
*
20 Hz.
8.
Record the measurement on the Performance Test Record Card, line 55.