Optimize pressure control
settings
To optimize pressure control parameters, change the settings manually and test the
new settings with a manual run. Follow the steps below:
Step
1
2
3
4
5
Set pressure control parameters
Follow the steps below to change the pressure control parameters in System:Settings.
Settings changed in this way will apply as default for both manual and method runs.
Step
1
ÄKTA pilot 600 User Manual 29274559 AA
Action
Open the Advanced:Pressure control parameters section of Manual instruc-
tions in System Control.
Adjust the parameters as required.
Activate pressure control and select the appropriate Pressure control pa-
rameter in the Pumps:System Flow section of Manual instructions.
Run Pumps:System flow at a suitable flow rate to test the effect of your
adjustments.
When you are satisfied, update the pressure control parameters in System
Settings so that the optimized settings will apply automatically in the future.
Action
Open System:Settings in the UNICORN System Control module.
4 Operation
4.5 Pressure control