GE DigitalFlow XGS868i Programming Manual page 14

Panametrics steam ultrasonic mass flow transmitter
Table of Contents

Advertisement

Chapter 1. Programming Site Data
1.3
The XGS868i Enclosure Keypad (cont.)
When you power up the XGS868i, the display first shows the model and software version:
The meter then starts to display measured parameters.
To enter the Keypad Program, press the [Escape] key, followed by the [Enter] key, and the [Escape] key again. Each
successive key must be entered within 10 seconds of the prior key.
As a guide in following the programming instructions in this chapter, the relevant portions of the Model XGS868i
menu map have been reproduced in Figure 28 on page 77 and Figure 29 on page 78. Proceed to the following sections
to enter data in the Channel or GLOBL menus.
IMPORTANT:
If the keypad has not been pressed for 10 minutes, the XGS868i exits the Keypad Program and returns to
displaying measurements. The meter retains any configuration changes that were confirmed with the
[Enter] key, and restarts as if the operator had completed the programming cycle.
4
GE
XGS868i Y4AS.STD
CH1
VEL
10.00
E1
Ft/s
DigitalFlow™ XGS868i Programming Manual

Advertisement

Table of Contents
loading

Table of Contents