GE DigitalFlow XGS868i Programming Manual page 117

Panametrics steam ultrasonic mass flow transmitter
Table of Contents

Advertisement

Transducer
Signal
Parameters
Range
Signal Low
20 to 100
Limit
Correlation
0 to 500
Peak Limit
Soundspeed
1 to 50%
+/– Limit
–500 to
Velocity
500 ft/sec
Low Limit
(–150 to
150 m/sec)
–500 to
Velocity
500 ft/sec
High Limit
(–150 to
150 m/sec)
Acceleration
0 to 250 ft/s
Limit
(0 to 76 m/s)
Amplitude
Discriminator
0 to 100
Low
Amplitude
Discriminator
0 to 100
High
-1000 to 1000
Delta T Offset
μsec
–500 to
Skan T Offset
500 μsec
% of Peak
1 to 100%
DigitalFlow™ XGS868i Programming Manual
Table 20: Transducer Signal Settings
Default
Value
The E1:LOW SIGNAL error message appears when the signal strength
20
falls below the programmed SIGNAL LOW LIMIT value. See Chapter 2 in
Service Manual
the
The E4: SIGNAL QUALITY error message appears when the signal quality
100
falls below the programmed COR. PEAK LIMIT value. See Chapter 2 in the
Service Manual
The E2:SOUNDSPEED error message appears when the calculated fluid
sound speed differs from the fluid sound speed entered in the Chan-
20%
nelx-System menu by more than the programmed SOUNDSPEED
± LIMIT value. See Chapter 2 in the
error codes.
The E3: VELOCITY RANGE error messages appears when the calculated
fluid velocity is less than the programmed VELOCITY LOW LIMIT value.
–150 ft/sec
See Chapter 2 in the
(–46 m/sec)
The E3: VELOCITY RANGE error messages appears when the calculated
fluid velocity exceeds the programmed VELOCITY HIGH LIMIT value. See
150 ft/sec
Chapter 2 in the
(46 m/sec)
The E6: CYCLE SKIP error message appears when the calculated fluid
15 ft/s
velocity changes by more than the programmed ACCELERATION LIMIT
value from one reading to the next. See Chapter 2 in the
(5 m/s)
ual
for a discussion of error codes.
The amplitude discriminator measures the transducer signal received
by the Model XGS868i. The default value for the above parameter is 14,
and values from 0 to 100 are acceptable. The E5: AMPLITUDE error
14
message appears when the amplitude discriminator falls below the
programmed AMP. DISCRIM LOW value. See Chapter 2 in the
Manual
for a discussion of error codes.
The amplitude discriminator measures the transducer signal received
by the Model XGS868i. The default value for the above parameter is 34,
and values from 0 to 100 are acceptable. The E5: AMPLITUDE error
34
message appears when the amplitude discriminator exceeds the
programmed AMP. DISCRIM HIGH value. See Chapter 2 in the
Manual
for a discussion of error codes.
An offset between the upstream and downstream transit times is speci-
0 μsec
fied at this prompt.
At this prompt, specify a time measurement offset that compensates for
58 μsec
any shift resulting from cross-correlation. Set to 0 for active Skan T
Offset.
The percentage of peak used to calculate the transit times and Delta T is
50%
specified at this prompt.
Appendix C. Programming the XGS868i Using PanaView™
Description
for a discussion of error codes.
for a discussion of error codes
Service Manual
Service Manual
for a discussion of error codes.
Service Manual
for a discussion of error codes.
for a discussion of
Service Man-
Service
Service
107