Ethylene Oxide Sterilization; Supplies - HP CodeMaster XL+ M1722A User Manual

Defibrillator/monitor
Table of Contents

Advertisement

Maintaining the Defibrillator

Supplies

Ethylene Oxide Sterilization

Temperature
Relative Humidity
EtO Concentration
Exposure Time
Aeration Time
Supplies
8
The Hewlett-Packard warranty is only assured if you use Hewlett-Packard approved
accessories and replacement parts.
Battery
M1758A
Patient Cables
Patient cable 6-pin/3-wire AHA
M1731A
Patient cable 5-wire/6-pin AHA
M1732A
Patient cable 8-pin/3-wire AHA
M1733A
Patient cable 5-wire/8-pin AHA
M1734A
Patient cable 8-pin/3-wire IEC
M1735A
Patient cable 5-wire/8-pin IEC
M1736A
Trunk cable 3-wire/12-pin - AHA
M1500A
Lead set 3-wire - AHA
M1605A
Trunk cable 3-wire/12-pin - IEC
M1510A
Lead set 3-wire - IEC
M1615A
Trunk cable 5-wire/12-pin - AHA
M1520A
Lead set 5-wire - AHA
M1625A
8-8
54ºC ± 1ºC (130ºF ± 2ºF)
60% ± 20%
600 mg/L ± 30 mg/L 12/88 EtO
1 hour 45 minutes
18 hours
Battery assembly