3
During defibrillator usage, the monitor stores up to 28 ECG strips of critical
information called events. Events include all shocks, heart rate alarms, SpO
and mark events. Each event record includes date of event, heart rate, ECG source,
and size setting as shown in Table 3-6, "Event Summary Record Information," on
page 3-8. The time annotated on the ECG strip is within 8 seconds of the recorded
event. The message "ES" is printed at the top of the ECG strip when you print the
Event Summary record.
Table 3-6

Event Summary Record Information

Event
Shock
Heart Rate
Alarms viola-
tion
SpO
violation
• To print the Event Summary on the recorder, press
• To stop printing the Event Summary, press
• To review the Event Summary later, turn the unit on and press
3-8
Event Summary Description
Shock#, Delivered energy, peak current, and patient impedance.
Heart Rate alarm limits.
Alarms
SpO
alarm limits.
2
2
Mark
Mark
Marker symbol (
not be printing to print an Event Summary with this key. After printing an event,
you must wait 10 seconds before printing another event.
Charge
Mark
) annotates strip at point
Review
Review
or
Record
.
alarms
2
was pressed.