Error Log - GE eBike III basic Series Service Manual

Electric bike with and without nibp
Table of Contents

Advertisement

e
rror log
The software of the eBike III ergometers stores internal
error messages in a special memory area (Error Logs) and
also stores statistics for the quality of the blood pressure
measurements (NIBP Logs).
These error messages can be checked and cleared in this
menu.
To display the stored error messages select „Show Error
Logs" and confirm with by pressing NIBP.
The first error message is displayed:
Source:
The unit generating the error
Err.Code:
Corresponding error code of the software
Quantity:
Number of occurences of this error
Using the arrow keys, the following error messages can be
displayed.
Press the NIBP (Exit) to close the error log displayed.
To display the bloodpressure statistics, select „Show NIBP
Logs" and confirm by pressing NIBP.
The statistics are displayed:
NIBP failed:
Number of failed measurements
Mic.Level 1...5:
Number of measurements with the
shown microphone signal strength
(5 = best signal)
2018111-336 Rev. B

Error log

Err. Log Quantity:
Show Error Logs
Show NIBP Logs
Clear Error Logs
Clear NIBP Logs
EXIT
e
rror log memory
Error No.:
Source:
Err.Code:
Quantity:
EXIT
d
iSplaying a Single error
NIBP failed:
Mic. level 1:
Mic. level 2:
Mic. level 3:
Mic. level 4:
Mic. level 5:
EXIT
e
rror log memory
eBike III basic / comfort
2
1 / 3
SMU
0100
1
2
2
5
3
9
7
nibp
- 67 -