11. Zero Calibration (Oxygen Span)

This must be performed in a clean atmosphere, and it is recommended that it be performed daily
or after any gas alarm. Depress the Test button for 5 seconds. The instrument will initiate a zero
by showing a countdown with a '0' for zero calibration, followed by a countdown from '20' to '0'.
When the zero calibration has been completed the XT will indicate a pass by beeping twice,
flashing twice, and the ( ) symbol will flash for 5 seconds.
blink
If the zero calibration is not OK then the XT will give a single beep and a single flash, the Test
Fail icon ( ) will be displayed, and the zero calibration should be repeated.