Entering Base (Zero) And Span Output Values; Entering Output Type And Units - GE UTX878 Ultrasonic User Manual

Ultrasonic flow transmitter
Table of Contents

Advertisement

March 2004
Entering Base (Zero)
and Span Output
Values
Entering Output Type
and Units
4-2
The 4-20 Loop submenu enables you to enter the information needed
to set up output parameters: unit type, base (zero) and span values,
and error handling. To enter data in the 4-20 Loop submenu:
1. Press [ESC]. The UTX878 enters the User Program.
2. Press the [ ] key until CONFIG is bracketed and press [ENT].
3. Press [ ] to reach the 4-20 Loop submenu, and press [ENT].
1. In the Loop submenu, press [ENT] to enter the Units option.
2. The menu displays five options: Channel 1, Channel 2, AVG,
SUM and DIFF. Use the [ ] and [ ] keys to scroll to the desired
option, and press [ENT].
Note: Channel 2, AVG, SUM and DIFF are only available for a two-
channel UTX878.
3. The screen now displays three measurement parameters: velocity,
volumetric, and soundspeed. Scroll to the desired selection and
press [ENT].
4. The menu now asks for the unit type.
For velocity or soundspeed measurements, the menu offers a
single selection (either feet/sec or meters/sec, depending on
your choice of units). Press [ENT] to confirm the entry and
return to the previous menu.
For volumetric, the menu offers a list of output units (either
metric or English). Scroll to the desired units and press [ENT].
Press [ESC] until you return to the CONFIG menu.
Configuring Meter Data