CHAPTER 1: PRODUCT DESCRIPTION

Testing and certification

MULTISYNC 100 GPS CLOCK – INSTRUCTION MANUAL
APPROVALS AND CERTIFICATION
Compliance
Applicable council directive
CE compliance
Low voltage directive
EMC directive
North America
cULus
ISO
Manufactured under a registered
quality program
IEC 61850-3 EMI TYPE TESTS
Test
Description
IEC 61000-4-2
ESD
IEC 61000-4-3
Radiated RFI
IEC 61000-4-4
Burst
IEC 61000-4-5
Surge
IEC 61000-4-6
Induced RFI
IEC 61000-4-8
Magnetic Field
IEC 61000-4-29
Voltage Dips and
IEC 61000-4-11
Interrupts
IEC 61000-4-12 Damped Oscillatory
IEC 61000-4-16 Mains Frequency
Voltage
IEC 61000-4-17 Ripple on D.C. Power
Supply
IEC 60255-5
Dielectric Strength
IEC 60255-5
H.V. Impulse
According to
EN60950-1
EN61000-6-2
EN61000-6-4
UL60950-1
C22.2 No. 60950-1
CB Report including all country deviations
ISO 9001:2008
Test Levels
Enclosure Contact +/- 8 kV
Enclosure Air
+/- 15 kV
Enclosure Ports
20 V/m
D.C. Power port
+/-4 kV
Signal Ports
+/- 4kV line to earth,
+/- 2kV line to line
D.C. Power Ports
+/- 2kV line to earth,
+/- 1kV line to line
Signal Ports
10 V
D.C Power ports
10 V
Earth Ground
10 V
Ports
Enclosure Ports
40 A/m continuous,
1000A/m for 1s
D.C. Power ports
30% for 0.1s, 60% for 0.1s,
100% for 0.05s
Signal Ports
2.5kV common, 1kV diff,
D.C. Power ports
mode @1MHz
Signal Ports
30V Continuous,
D.C. Power ports
300V for 1s
D.C. Power ports
10%
Signal Ports
2 kVAC
(fail-safe relay output)
D.C. Power ports
2 kVAC
Signal Ports
5 kV (fail-safe relay output)
SPECIFICATIONS
Severity
Levels
4
4
4
4
3
3
3
3
3
4
3
5