Hanna Instruments HI 8614 Instruction Manual

Ph and orp transmitters
Hide thumbs Also See for HI 8614:

Advertisement

Instruction Manual
HI 8614 • HI 8614L
HI 8615 • HI 8615L
pH and ORP Transmitters
p H
L
H I 8 6 1 4
C A T O R
p H I N D I
R
S M I T T E
& T R A N
w w w . h a n n a i n s t . c o m
1

Advertisement

Table of Contents
loading

Summary of Contents for Hanna Instruments HI 8614

  • Page 1 Instruction Manual HI 8614 • HI 8614L HI 8615 • HI 8615L pH and ORP Transmitters H I 8 6 1 4 C A T O R p H I N D I S M I T T E & T R A N w w w .
  • Page 2: Table Of Contents

    Dear Customer, PRELIMINARY EXAMINATION Thank you for choosing a Hanna Instruments Product. Remove the instrument from the packing ma- Please read this instruction manual carefully terial and examine it carefully to make sure before using the instrument. that no damage has occurred during ship- ping.
  • Page 3: Specifications Of Hi 8614 & Hi 8614L

    The transmitters use a universal BNC socket SPECIFICATIONS OF HI 8614 & HI 8614L for quick and secure connection to any elec- trode with a BNC connector. Range 4 to 20 mA / 0.00 to 14.00 pH For HI 8614 and HI 8614L the temperature Resolution 0.01 mA / 0.01 pH compensation is performed by the transmitter's...
  • Page 4: Specifications Of Hi 8615 & Hi 8615L

    SPECIFICATIONS OF HI8615 & HI8615L TERMINAL BOARD CONNECTIONS Unscrew the 4 screws and remove the top cover. Range 4 to 20 mA / ±1000 mV Resolution 0.01 mA / 1 mV There are three cable glands on the transmitter Accuracy (@20°C/68°F) cover.
  • Page 5: Ph Calibration With Automatic Temperature Compensation (Hi8614 & Hi8614L)

    The wire between the transmitter and the pH CALIBRATION WITH AUTOMATIC recorder/indicator/controller should be a PVC TEMPERATURE COMPENSATION insulated two wire with a wire diameter of at (HI 8614 & HI 8614L) least 0.7 mm. This wire is fed through one of the smaller cable glands.
  • Page 6 Procedure For other buffer temperatures, refer to page 13 for the appropriate mA / pH reading. • Disconnect the +ve supply cable from "+VE LOOP" terminal and connect a 20mA • Rinse the electrode with tap f.s. ammeter between the +ve cable and water or distilled water and "+VE LOOP"...
  • Page 7 Under normal application, adjustment on this pH CALIBRATION WITH FIXED module may not be necessary. TEMPERATURE COMPENSATION If routine check is required, the following pro- (HI 8614 & HI 8614L only) cedures shall be performed. • Take the temperature of the •...
  • Page 8: Orp Calibration (Hi8615 & Hi8615L)

    • If the reading lies outside ORP CALIBRATION (HI8615 & HI8615L) this range, adjust the slope SLOPE ADJUSTMENT adjustment trimmer on the Initial preparation OFFSET ADJUSTMENT transmitter for a reading just Disconnect the +ve supply cable from the within this range. "+VE LOOP"...
  • Page 9 • Set the simulator to 350 mV and adjust FOR HI 8615L ONLY the slope trimmer to read 14.8 mA on the The HI 8615L is factory calibrated, and the ammeter or 350 mV on the HI 8615L displayed values are referenced to the 4-20 display (HI 8615L only).
  • Page 10: Electrode Conditioning And Maintenance

    Please note: ELECTRODE CONDITIONING AND MAINTENANCE -1000 mV 4 mA 0 mV 12 mA 1000 mV 20 mA Reference 350 mV 14.8 mA Filling Hole 200 mV 13.6 mA Reference 275 mV 14.2 mA Filling Hole 1 mV 12.008 mA Note: when the meter is used in conjunction Sensitive Reference...
  • Page 11 "shaking down" the electrode as you would PERIODIC MAINTENANCE do with a glass thermometer. Inspect the electrode and the cable. The cable If the bulb and/or junction are dry, soak the used for connection to the meter must be electrode in HI70300 storage solution for at intact and there must be no points of broken least one hour.
  • Page 12: Temperature-Resistance Correlation For Hanna Ph Sensitive Glass

    TROUBLESHOOTING TEMPERATURE-RESISTANCE Evaluate your electrode performance based CORRELATION FOR HANNA on the following. pH SENSITIVE GLASS • Noise (Readings fluctuate up and down) The resistance of glass electrodes partially could be due to: depends on the temperature. The lower the - Clogged/Dirty Junction: refer to the temperature, the higher the resistance.
  • Page 13: Installation Procedure And Examples

    INSTALLATION PROCEDURE Typical Electrode Life AND EXAMPLES Ambient Temperature 1- 3 years 90 °C Less than 4 months 120°C Less than 1 month The HI 8614, HI 8614L, HI 8615 and HI 8615L transmitters may be wall mounted in any con- venient location near the measurement point.
  • Page 14 Monitoring the pH/ORP with Panel Mounting Monitoring and Controlling the pH with pH (HI 8510)/ORP (HI 8512) Indicator (HI 8711) Panel Mounting Indicator/ Regulator with Independent Dosage Control for Acid and Base Controlling the pH/ORP with an Industrial Regulator Monitoring and Controlling the pH/ORP with Panel Mounting Indicator/Regulator and Dosage Control of either Acid or Base...
  • Page 15: Accessories

    pH ELECTRODES ACCESSORIES HI 1090T Screwcap PG13.5 connector, pH CALIBRATION SOLUTIONS double junction, glass body HI 7004M pH 4.01 buffer solution, 230 mL PG13.5 THREAD φ 12mm φ 9.5mm HI 7004L pH 4.01 buffer solution, 500 mL HI 7006M pH 6.86 buffer solution, 230 mL HI 7006L pH 6.86 buffer solution, 500 mL HI 7007M pH 7.01 buffer solution, 230 mL 30mm...
  • Page 16 HI 1210T Screwcap PG13.5 connector, HI 2910B/5 BNC connector, 5 m (16.5') cable, double junction, plastic body double junction, plastic body, built- in amplifier PG13.5 THREAD φ 12mm 3/4 x 16 UNF DIA 12mm DIA 20.5mm 38.5mm 110mm 30mm 110mm HI 1910B BNC connector, 1 m (3.3') cable, ORP ELECTRODES...
  • Page 17 HI 3210T Screwcap PG13.5 connector, Pt, EXTENSION CABLES FOR SCREW-TYPE plastic body ELECTRODES ONLY (SCREW TO BNC CON- NECTOR) PG13.5 THREAD φ 12mm HI 7855 SERIES CABLE CONNECTORS CONNECTOR AND 3.0 mm (0.12") CABLE WITH BNC CONNECT TO THE CONNECT TO BNC SOCKET SCREW TYPE OF THE METER...
  • Page 18: Warranty

    WARRANTY CE DECLARATION OF CONFORMITY All Hanna Instruments meters are warranted for two years against defects in workman- ship and materials when used for their in- tended purpose and maintained according to instructions. The probes and the electrodes are war- ranted for a period of six months.
  • Page 19 SALES AND TECHNICAL SERVICE CONTACTS Australia: Tel. (03) 9769.0666 • Fax (03) 9769.0699 China: Tel. (10) 88570068 • Fax (10) 88570060 Egypt: Tel. & Fax (02) 2758.683 Germany: Tel. (07851) 9129-0 • Fax (07851) 9129-99 Greece: Tel. (210) 823.5192 • Fax (210) 884.0210 Indonesia: Tel.

This manual is also suitable for:

Hi 8615Hi 8614lHi 8615l

Table of Contents