2-6
CHARACTERISTICS
Min. trigger sensitivity:
d.c. to 1 MHz
via EXT input
d.c. to 40 MHz
via EXT input
TV (ampl. of sync. pulse)
via EXT input
SLOPE selection:
LEVEL control range
SPECIFICATION
1 div
150 mV
2 div
300 mV
1 div
100 mV
positive going (+)
negative going (—)
+7 div
2.9
BLANKING
OR
Z AXIS
Input coupling
Input voltage limits
Sensitivity
2.10
CALIBRATOR
Wave form-shape
Output voltage (p-p)
Frequency
d.c.
+42V
3 V (p-p)
square-wave
1V
1 kHz
2.11
POWER
SUPPLY
Source voltage a.c. (r.m.s.):
nominal
limits of operation
Source frequency:
nominal
limits of operation
Power consumption:
typical
maximum
100, 120, 220, 240 V
Nominal voltage + 10 %
50...400 Hz
48...440 Hz
35 W
40 W
2 CHARACTERISTICS
ADDITIONAL INFORMATION
Trigger level not in fixed position
Trigger level not in fixed position
Trigger level not in fixed position
Trigger !evel not in fixed position
Positive and negative video,
selection via SLOPE switch
Positive and negative video,
selection via SLOPE switch
Typical
At max. 1 kHz
Pos. signal blanks display
Factory installed
At nominal source voltage