Advertisement
This Monitor is designed for adults and should never be used on infants, young children or patients with preeclampsia.
This product might not meet its performance speci cations if stored or used outside the speci ed temperature and humidity ranges.
Please do not share the cuff with any infectious person to avoid cross-infection.
Blood Pressure Monitor that uses the oscillometric principle to measure your blood pressure and pulse rate. Blood pressure fluctuates continually day and night. Any physical activity or excitement will influence your blood pressure. Because of this, it is unusual to obtain identical multiple blood pressure readings. This is why it is recommended that you measure your blood pressure at approximately the same time every day.
The explanation of symbols on the display:
Bluetooth Symbol
Synchronization Symbol
Release Air Symbol
Memory Indicator
Low Battery Indicator
Irregular Heartbeat Symbol
Blood Pressure Level Classification Indicator
Taking a measurement is easy. It's what you do with the measurement that is important. You can use your device as a stand-alone device just to measure your blood pressure, but when you connect it to the free iHealth MyVitals App you will be able to track, store, manage and share your results with your family, doctor or care givers. Simply download iHealth MyVitals on the Apple Store or on GooglePlay and start managing your health.
Your iHealth account also gives you access to the free and secure iHealth cloud service so that you can manage your data, watch your trends and share your data. To do this, visit www.ihealthlabs.eu and log in using the same account as you used with your App.
The iHeath Track Blood Pressure Monitor is designed to be used with Bluetooth 4.0 Apple devices and selected Bluetooth 4.0 Android smartphones. Please note that the compatible devices are subject to change. For the latest compatibility list visit www.ihealthlabs.eu/support.
When the display shows the battery symbol ( ), replace all the batteries with new ones. Rechargeable batteries are not suitable for this monitor.
The iHeath Track Blood Pressure Monitor has a universal size cuff. This means that it will fit any arm with a circumference between 22 and 42cm. If your arm is too large to fit in this cuff, you might consider using a wrist blood pressure monitor.
Note: If the cuff becomes dirty, remove it from the monitor and wipe it with a lightly moistened soft cloth. Then air-dry the cuff. Do not machine wash. Cleaning the cuff after every 200 measurements is recommended.
Once you have determined that the cuff fits your arm (22 – 42cm), simply insert the cuff connector into the socket on the side of the monitor. Ensure that the connector is completely inserted to avoid air leakage.
Avoid compression or restriction of the connection tubing during measurement which may cause in ation error, or harmful injury due to continuous cuff pressure.
Insert the batteries in the device and connect it with the iHealth MyVitals App on your smartphone. The time and date will automatically be updated. If you change the batteries, please readjust the time and date again.
CLASSIFICATION OF BLOOD PRESSURE AND BACKLIGHT INDICATOR | |||
BLOOD PRESSURE CLASSIFICATION | Systolic mmHg | Diastolic mmHg | BACKLIGHT INDICATOR |
Optimal | 91–119 | and 61 - 79 | Green |
Normal | 120-129 | and 80 -84 | Green |
High-normal | 130-139 | or 85 - 89 | Yellow |
Grade 1 Hypertension | 140-159 | or 90 - 99 | Flashing Yellow |
Grade 2 Hypertension | 160-179 | or 100–109 | Red |
Grade 3 Hypertension | ≥180 | or ≥110 | Flashing red |
You can review previous readings by pressing the "M" button twice. The rst memory displayed will be the average of the last three measurements. When the average of the last three measurements display, press the "M" button, the newest measurement will display. Press "M" button again to review the next measurement. During display the memories, keep on pressing button "M" for three seconds, all results in the current memory bank will be deleted.
If no result stored, LCD will show zero.
The monitor will automatically switch off after one minute to save the batteries. Alternatively, you can press the "START/STOP" button to turn off the monitor manually.
Before you can synchronize all your blood pressure readings, you need to download the iHealth MyVitals App on your smartphone from the Apple Store (for iOS devices) or on GooglePlay Store (for Android). You then need to create your own iHealth account on the App.
To synchronize:
Problem | Explanation | Solution |
LCD shows low battery symbol | Low Battery | Change all the batteries |
LCD shows "Er 0" | Pressure system is unstable before measuremente | Try again without moving |
LCD shows "Er 1" | Fail to detect systolic pressure | |
LCD shows "Er 2" | Fail to detect diastolic pressure | |
LCD shows "Er 3" | Pneumatic system blocked or cuff is too tight during inflation | Apply the cuff correctly and try again |
LCD shows "Er 4" | Pneumatic system leakage or cuff is too loose during inflation | |
LCD shows "Er 5" | Cuff pressure above 300mmHg | Measure again after 5 minutes. If the error is still there, please contact your local iHealth distributor |
LCD shows "Er 6" | More than 3 minutes with cuff pressure above 15 mmHg | |
LCD shows "Er 7" | EEPROM accessing error | |
LCD shows "Er 8" | Device parameter checking error | |
LCD shows "Er A" | Pressure sensor parameter erroron | |
No response when you press button or install batterypiles | Incorrect operation or strong electromagnetic interference. | Take out the batteries, wait five minutes and then reinstall batteries |
Please visit the support section of the www.ihealthlabs.eu website to see our comprehensive Frequently Asked Questions on blood pressure.
If you have any questions regarding the aftersales or usage of the product, please visit www.ihealthlabs.eu and consult our Frequently Asked Questions on blood pressure (in our support section). If you do not nd the answers you are looking for, please send us a query directly from the website. If you have any suggestions so that we can improve our products, please send us a mail at support@ihealthlabs.eu.
ANDON HEALTH CO., LTD
Add: No. 3 Jinping Street, YaAn Road, Nankai District, Tianjin 300190, China.
Tel: 86-22-60526161
Product name: iHeath Track Blood Pressure Monitor
Model: KN-550BT
Classification: Internally powered, type BF applied part; IPX0, No AP or APG; Continuous operation
Machine size: 98 × 98 × 46mm
Cuff circumference: 22 - 42cm
Weight: 348g (batteries and cuff included)
Measuring method: oscillometric method, automatic inflation and measurement
Memories: 99
Power source: 4x1.5V SIZE AAA batteries
Measurement range:
Cuff pressure: 0 - 300mmHg
Systolic: 60 - 260mmHg
Diastolic: 40 - 199mmHg
Pulse rate: 40 - 180 beats/minute
Measuring accuracy: Pressure: ±3mmHg, Pulse rate: ±5%
Environmental temperature for operation: 10~40°C
Environmental humidity for operation: ≤85%
Environmental temperature for storage: -20~50°C
Environmental humidity for storage: ≤85%
Environmental pressure: 80~105kPa
Battery life: Approximately 250 readings
This blood pressure measurement system includes accessories: pump, valve, cuff, and sensor.
Note: These specifications are subject to change without notice.