Download Print this page
Honeywell AutoRAE 2 Quick Start Manual
Honeywell AutoRAE 2 Quick Start Manual

Honeywell AutoRAE 2 Quick Start Manual

Hide thumbs Also See for AutoRAE 2:

Advertisement

Quick Links

by Honeywell
Setup
Before using the AutoRAE 2 Controller and AutoRAE 2 Cradle(s) to bump test or calibrate ToxiRAE Pro-family,
QRAE 3, MicroRAE, Handheld PID, and/or MultiRAE-family (pumped versions) instruments, follow the
instructions in the AutoRAE 2 User's Guide for programming the gas configurations. The steps to follow for
assembly and gas configuration, as well as functionality, are the same for all cradles.
Connections
Assemble the system. Connect the AutoRAE 2 Controller to any Cradles, and then connect the Terminator. For
ToxiRAE Pro Cradles, insert the appropriate adapter.
Attach Controller to Cradle(s) and then
attach Terminal Adapter to last Cradle.
Exhaust
Calibration gas cylinders
(each must have a demand-
flow regulator)
Important! Make sure an SD card is installed in the AutoRAE 2 Controller. The system cannot operate if an SD
card is not installed. For installation instructions, consult the AutoRAE 2 User's Guide. Make sure the AutoRAE 2
Controller's RTC (real-time clock) has been set in Safety Suite Device Configurator (SSDC) before its first use.
AutoRAE 2
AutoRAE 2
Controller
Cradles
(maximum
of 10)
www.sps.honeywell.com
AutoRAE 2 System
Quick Start Guide
Insert and tighten
screws, and then put
caps over screws.
Press filter onto
receptacle at end
of Controller
UltraRAE 3000 or ppbRAE 3000 only:
!
Use an Active Carbon
Filter for VOC filtering
The ToxiRAE Pro Cradle requires
an adapter to be inserted in the
capture mechanism. ToxiRAE Pro
EC (L) and ToxiRAE Pro LEL (L)
and ToxiRAE Pro PID (R).
Press filter onto
receptacle at
end of Cradle
Plug the barrel from AC adapter into
jack. Then plug its cord into the
transformer, and the transformer
into AC power source.

Advertisement

loading

Summary of Contents for Honeywell AutoRAE 2

  • Page 1 AC power source. Important! Make sure an SD card is installed in the AutoRAE 2 Controller. The system cannot operate if an SD card is not installed. For installation instructions, consult the AutoRAE 2 User’s Guide. Make sure the AutoRAE 2 Controller’s RTC (real-time clock) has been set in Safety Suite Device Configurator (SSDC) before its first use.
  • Page 2: Turning The System On

    The AutoRAE 2’s internal pump and the valves in the system are tested. If self-checkout is successful, the system is ready to use. The display provides a list of the AutoRAE 2 Cradles, and if there are instruments in them, the names of the instruments are included in this list.
  • Page 3: Bump Testing

    AutoRAE 2 System Quick Start Guide by Honeywell Bump Testing You can bump test all or individually selected instruments that are seated in the system’s cradles. From the main screen, enter Bump Test by pressing [Y/+] twice (“Function” and then “Select”).
  • Page 4 AutoRAE 2 System Quick Start Guide by Honeywell Calibration You can calibrate all or individually selected instruments that are seated in the system’s cradles. From the main screen, enter Calibrate by pressing [Y/+] (“Function”), and then [N/-] until “Calibrate” is highlighted.
  • Page 5 MicroRAE Charging An Instrument’s Battery Each AutoRAE 2 Cradle attached to an AutoRAE 2 Controller automatically charges an instrument’s battery when it is docked. The Charge Status LED glows red to indicate that charging is taking place. When the instrument’s battery is fully charged, the LED glows green.
  • Page 6: Corporate Headquarters

    UltraRAE 3000: MiniRAE Lite, MiniRAE 3000, or ppbRAE 3000: Do not use a RAE-Sep Tube in the UltraRAE 3000 tube holder when using it with an AutoRAE 2 Cradle. CORPORATE HEADQUARTERS WORLDWIDE SALES OFFICES RAE Systems by Honeywell USA/Canada: 1.877.723.2878 300 S Tryon St.