Subsystem Syntax

CLOSe

Parameters
Comments
Example
58 Relay Multiplexer Command Reference
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The [
subsystem controls switching and scanning operations for the
ROUTe:]
multiplexer modules in a switchbox.

[ROUTe:]

CLOSe <
channel_list
CLOSe? <
channel_list
OPEN <
channel_list
OPEN? <
channel_list
SCAN <
channel_list
:MODE NONE|VOLT|RES|FRES
:MODE?
:PORT ABUS|NONE
[ROUTe:]CLOSe <
channel_list
channel_list. The channel_list is in the form (@ccnn), (@ccnn,ccnn), or
(@ccnn:ccnn) where cc = card number (00–99) and nn = channel number (00–15).
Parameter Name
<
channel_list
>
Closing Channels: To close:
-- a single channel, use
-- multiple channels, use
-- sequential channels, use
-- a group of sequential channels, use
[ROUTe:]CLOSe (@ccnn:ccnn,ccnn:ccnn)
-- or any combination.
Closure order for multiple channels with a single command is not guaranteed.
Closing the AT, BT, AT2, and RT Tree Switches: Use channel numbers 90,
91, 92, and 93 to close the AT, BT, AT2, and RT Tree Switches, respectively.
Related Commands:
*RST Condition: All multiplexer channels are open.
Closing Multiplexer Channels
This example closes channel 00 of a card number 1 multiplexer module and
channel 15 of a card number 2 multiplexer module in a single switchbox.
CLOS (@100,215)
>
>
>
>
>
>
closes multiplexer channels specified in the
Parameter Type
numeric
[ROUTe:]CLOSe (@ccnn)
[ROUTe:]CLOSe (@ccnn,ccnn)
[ROUTe:]CLOSe (@ccnn:ccnn)
;
[ROUTe:]OPEN
,
[ROUTe:]CLOSe?
100 closes channel 00 of multiplexer #1;
215 closes channel 15 of multiplexer #2.
[ROUTe:]
Range of Values
cc00-cc15
;
;
;