(2) Wired connection (option)
No.
Item
Installing the driver of
1
infrared data
communication unit,
Connecting the infrared
data communication unit
2
with the image capture
computer
Turning on the power of
3
the detector
4
Registration of detector
Connecting the detector
5
and AC/DC box using the
sensor cable
Connecting the status
6
indicator to the AC/DC
box
Connecting the image
7
capture computer with
the AC/DC box
Turning on the power of
8
the AC/DC box
Connecting the image
9
capture computer with
the X-ray interface box
Connecting the X-ray
10
generator to the X-ray
interface box
Establishing the CXDI
11
Control Software NE
settings
Connecting to the
12
network and setting the
output destination
Aligning the detector and
13
X-ray tube exposure
fields
14
Calibration
15
Self-diagnosis test
16
Imaging test
17
Final check
18
Cleaning
* When performing both wireless connections and wired connections for one detector, first proceed
with installation items No.1 to No.5 for the wireless connections.
2. Installation procedures
Conditions and checkpoints
• The driver installation must be completed
successfully.
• Connect the unit to the USB port on the image capture
computer.
• Turn on the power of the image capture computer and
start the link controller.
• A fully charged battery must be installed properly in
the detector.
• The LED on the detector must light.
• The distance between the detector and infrared data
communication unit must be less than 30 cm.
• Care must be taken not to drop the detector.
• The detector must be registered in a location outside
the patient environment.
• Check the directions in which the cable connectors
are inserted, and plug them in.
• Check the directions in which the cable connectors
are inserted, and plug them in.
• This connection must be made through a hub.
• Route the LAN cable in such a way that no excessive
weight will be applied to it.
• The battery must be installed properly in the detector.
• The LED on the detector must light.
• This connection must be made through a hub.
• Route the LAN cable in such a way that no excessive
weight will be applied to it.
• Route the X-ray interface cable in such a way that no
excessive weight will be applied to it.
• Set the items which are required.
• The center positions of the detector and X-ray tube
must be aligned.
• No errors must be displayed.
• No errors must be displayed.
• Transfer images to the printer and storage unit, and
check their image quality.
• Carry out the final checks as per the check sheet.
- 19 -
Reference items
Refer to "7.1.2
Connecting the
infrared data
communication unit".
Refer to "7.1.2
Connecting the
infrared data
communication unit".
Refer to "7.1.6
Registration of
detector".
Refer to "7.1.6
Registration of
detector".
Refer to "7.1.8 When
connecting the wiring
unit".
Refer to "7.1.8 When
connecting the wiring
unit".
Refer to "7.1.8 When
connecting the wiring
unit".
Refer to "7.1.7
Connecting to the
X-ray interface box".
Refer to "7.1.7
Connecting to the
X-ray interface box".
Refer to "7.4 CXDI
Control Software NE
settings".
Refer to "7.4 CXDI
Control Software NE
settings".
Refer to "7.5 Image
Quality Checks".
Refer to "3.4 QC Tool
Button" in CXDI
Control Software NE
Setup guide.
Refer to "3.4 QC Tool
Button" in CXDI
Control Software NE
Setup guide.
Refer to "7.5 Image
Quality Checks".
Refer to "7.6
Post-installation
checks".