User Manual
2.11 Example Procedure: Measure/Simulate RTD
The following examples describe how to measure or
simulate an RTD.
Note: The same configurations are used to measure or
simulate resistance Ω. In each case, at the
DPI620G-IS, select the Resistance function (Range 0
to 4000 Ω) rather than the RTD function.
Note: A 4-wire configuration provides the best accuracy, a
2-wire configuration the lowest accuracy.
2.11.1
Example Procedure: Measure/Simulate a RTD, 4-Wire
Method
Figure 2-23
or simulate an RTD using the 4 wire method.
Figure 2-23 PT100 RTD Measure CH1 4-Wire
(Range -200 to 850°C)
29 July 2016
and
Figure 2-24
116M5464 Revision -
show CH1 set-up to measure
Druck DPI620G-IS
Page 64 of 248