MHR/P Source

Revision C
Maternal Heart/Pulse Rate Monitoring: MHR/P Source
The MHR/P can be determined by the MECG, MSpO
monitor. However, the data from only one parameter is:
referred to as the MHR/P source
displayed in the MHR/P area;
used to evaluate an MHR/P alarm condition; and
used to generate the MHR/P trace on the strip chart paper.
The source is:
selected via the MHR/P Setup screen
may be manually selected or automatically selected by the monitor according to
the following priority order:
MECG (values updated continuously)
MSpO
(values updated continuously)
2
NIBP (NIBP is available only as an Auto selection; the Manual selection is
disabled. Values updated only when NIBP determinations are taken)
IMPORTANT:
MSpO
AS AN MHR/P SOURCE—If MSpO
2
MHR/P source, an MHR/P alarm only occurs if the pulse rate
value derived from the MSpO
limit. The MHR/P values derived from the MECG and NIBP
sections of the monitor are ignored. The heart rate tone varies in
pitch to reflect changes in the maternal oxygen saturation reading.
The pitch rises as the saturation values increase, and lowers as the
saturation values decrease. The pulse rate trend is a grey line
annotated by
MSpO
1
If NIBP is selected as the MHR/P source, there is no trending of the data since these are static
measurements.
250cx Series Maternal/Fetal Monitor
2036946-001
, and NIBP sections of the
2
2
sensor violates an MHR/P alarm
2
P
.
2
1
is selected as the
7-3