NO SYNC/SYNC OPTION

The Responder 2000 has a No Sync/Sync Option available in Manual Mode only. If No Sync is selected, shock is
delivered immediately. When Sync is selected, full length R-wave sync marks are drawn on display and shock delivery is
synchronized with the R-wave. No Sync is selected by default.
1. Turn the Rotary Selector Knob to No Sync/Sync toggle button
2. Press the Rotary Selector Knob to select Sync option
PRECAUTION: Observe the ECG rhythm. Confirm that the full length sync bar appears near the middle of each QRS
complex. If the sync bars do not appear or are displayed in the wrong locations change the lead source.
NOTE: Every time after a Sync Shock is delivered; the device resets to "No Sync".
NOTE: Select a lead that provides a unipolar R-wave with a minimum amplitude of 1 mV and a low-amplitude T-
wave.

AUTO SEQUENCE

In Manual Mode, there is also an Auto Sequence option. In Auto Sequence, three energy levels can be pre-set. When
this option is selected, the Responder 2000 automatically selects the first pre-set energy and prompts the operator to
push the Charge button. After the Responder 2000 is charged, it will prompt the operator to push the Shock button(s).
After a shock is delivered, the Responder 2000 automatically selects the next energy level. After all three pre-set shocks
have been delivered; the Responder 2000 selects the last energy level. All shocks after the third shock will be at that
level.
NOTE: In Auto Sequence if the user wants to deliver a Sync Shock, Sync must be selected individually for all
shocks in Auto Sequence.