Verification

26
Verification
Using an ECG stored on a floppy disk, transmit a record to the MUSE
CV system.
1. Turn on the MAC 5000.
2. Insert a floppy disk with an patient ECG.
127(
If no ECG is available for the verification, take a flat line ECG
with no patient ID number and store it to a floppy disk.
3. Press the File Manager button.
4. Select the ECG you would like to transmit.
5. Press the Transmit button.
6. Verify the selected ECG transmitted to the MUSE CV system.
7. Repeat steps 1 through 8 for each MAC 5000 configured with a client
bridge.