4.6.3 Pulse oximetry

A pulse oximeter measures the light absorption of blood at two wavelengths, one in the near
infrared (about 940 nm) and the other in the red region (about 660 nm) of the light spectrum.
These wavelengths are emitted by LEDs in the SpO
peripheral tissue and is finally detected by a PIN-diode opposite the LEDs in the probe. The
pulse oximeter derives the oxygen saturation (SpO
relationship between the relative absorption at the two wavelengths and the arterial oxygen
saturation SaO
In order to measure the arterial saturation accurately, pulse oximeters use the component of
light absorption giving variations synchronous with heart beat as primary information on the
arterial saturation.
A general limitation of pulse oximetry is that due to the use of only two wavelengths, only two
hemoglobin species can be discriminated by the measurement.
The modern pulse oximeters are empirically calibrated either against fractional saturation
SaO
frac;
2
SaO
2
or against functional saturation SaO
SaO
2
Functional saturation is more insensitive to changes of carboxyhemoglobin and
methemoglobin concentrations in blood.
The oxygen saturation percentage SpO
against functional saturation SaO
SpO
measurement relative to SaO
2
of carboxyhemoglobin in blood. Independent of the calibration method, pulse oximeters are
not able to correctly measure oxygen content of the arterial blood at elevated
carboxyhemoglobin or methemoglobin levels.
Plethysmographic pulse wave
The plethysmographic waveform is derived from the IR signal and reflects the blood pulsation
at the measuring site. Thus the amplitude of the waveform represents the perfusion.
1-44
.
Document no
2044677-001
.
2
HbO
=
frac
+
+
HbO
Hb
2
HbO
2
=
func
+
HbO
Hb
2
probe, the light is transmitted through
2
using an empirically determined
)
2
2
Dyshemoglo
bin
func;
2
measured by the Datex-Ohmeda module is calibrated
2
func. The advantage of this method is that the accuracy of
2
func can be maintained even at rather high concentrations
2
Formula 1
Formula 2