Honeywell Signalpoint Pro Technical Manual page 25

Table of Contents

Advertisement

SIGNALPOINT PRO OPERATING MANUAL
9. When the gas reading is stabilised, exit the test mode by swiping the magnet
once over the oval mark during the test mode. The instrument then displays "
rESEt " for 15 seconds.
10. Wait for the instrument to time out of the " rESEt " display mode and the
instrument then enters the calibration mode.
11. Recalibrate the detector (see step 3 of section 8.3.1 for toxic version or step 3
of section 8.3.2 for Oxygen version).
Diagram 7: Replacing plug in sensor
24
MAN0853 Issue 3 10/08
Sensor Cover
Sensor retaining ring
Sensor
Transmitter

Advertisement

Table of Contents
loading

Table of Contents