3.14 Switching over the program for interpretation

Switching over the program for interpretation is activated by pressing the "H" key, the
following display image appears:
I n t e r p r e t a t i o n
1=
HEART
2=
12SL
X=
End
The currently enabled interpretation program appears in reverse video.
Warning!
Switching over to another program for Interpretation affects the measurement
results and the interpretation!
The program for Interpretation configured during manufacture should only be
changed when deemed really necessary!
The required program for Interpretation is selected by pressing the appropriate key,
"1" or "2".
Quit the selection menu by pressing the "X" key.
Quit the self-test with the "X" key.
MAC 1200 configures the appropriate items and initiates a cold start automatically.
Thus, when the device is rebooted automatically all the new settings are adopted.
Note on saved ECGs when switching over the program for Interpretation:
Saved ECGs are not lost.
After switching over to a different program for Interpretation when printing the saved
ECGs the display in the status line (12SL, or without in the case of HEART) is related
to the newly selected interpretation program, although the results are based on the
program previously configured. In this case the saved ECGs should be printed out or
transferred to a PC before switching over to the new program for Interpretation.
MAC 1100/1200 V 1.1
227 492 20 D - 0002
Page 38