Cycles cannot be turned on if the tank water level does not reach the minimum.
The autoclave is set with two rear automatic loading attachments for water, to be used
with a water purifier or external can, using supplied fittings. In this case, accessories
configuration and installation must be performed by an authorised technician.
In case of serious emergencies - loading system malfunctions and sterilization
requirements - it is possible to top up the tank via the auxiliary input located under the top
level (sheet metal plate cover, as seen in the following image).
Report the loading system malfunction to Smeg technical assistance before
carrying out the procedure described below.
1. Switch off the machine and disconnect the power supply cable connected at the
rear of the equipment.
2. Remove the sheet metal cover, holding it at the rear and pulling upward. The cover
is fastened with inserts which lock and unlock when pressed.
housed under the cover. These tanks have blue rubber covers, do not remove
them.
3. A white plastic cap, 4 cm diameter, identifies the auxiliary emergency input.
4. Remove the white cap and slowly pour demineralised water into the hole (max 2
litres), taking care that none leaks (the electrical and electronic components on the
machine could become damaged)
5. Reposition the white plastic cap
6. Reposition the sheet metal cover.
fig. 19 – Top cover removed for auxiliary water input use - only for emergencies
Revision n.1