Philips HEARTSTART XLT Service Manual page 126

Advertisement

Top Case Assemblies
ECG Connector
Interface PCA
4-40
ECG Connector
The ECG connector is located in the corner of the top case, next to the SpO
connector.
Removal
1. Disconnect and remove the Control Stack. See "Control Stack" on page
4-33.
2. Remove the plastic insulating shield over the SpO
Remove the tape that secures the shield to the metal clips around the SpO
and ECG connectors and pull the shield straight up.
3. Referring to Figure 4-22, disconnect the cable from the ECG connector to
the Interface PCA at the Interface PCA. Press the latch on top of the con-
nector and pull straight out. Swing the cable up out of way and lift it off
the case post.
4. Release the locking tabs on the metal clip around the ECG connector
using a small flat-bladed screwdriver. Pull up and remove the metal clip.
5. Slide the connector, cable, gasket and ferrite out through the hole in the
case.
Figure 4-24 Removing the ECG Connector
Metal Clip
and Interface PCAs.
2
Cable
Ferrite
Removal and Replacement
2
2