Safety Information
Guidance and Manufacturer's Declaration – Electromagnetic Emissions
The CXDI-50G is intended for use in the electromagnetic environment specified below.
The user of the CXDI-50G should assure that it is used in such an environment.
Emission Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Immunity Test
Electrostatic dis-
charge (ESD)
IEC 61000-4-2
Electrical fast tran-
sient/burst
IEC 61000-4-5
Surge
IEC 61000-4-5
Compliance
GROUP 1
The CXDI-50G uses RF energy only for its internal func-
tion. Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electromagnetic
equipment.
Class A
The CXDI-50G is suitable for use in all establishments
other than domestic and those directly connected to the
public low-voltage power supply network that supplies
Not Applica-
buildings used for domestic purposes.
ble
Complies
IEC 60601
Compliance
Test Level
Level
±(2, 4, 6) kV
±(2, 4, 6) kV
contact
contact
±(2, 4, 8) kV
±(2, 4, 8) kV
air
air
±2 kV for
±2 kV for
power supply
power supply
lines
lines
±1 kV for
±1 kV for
input/output
input/output
lines
lines
±1 kV differ-
±1 kV differ-
ential mode
ential mode
±2 kV com-
±2 kV com-
mon mode
mon mode
Electromagnetic Environment – Guidance
Electromagnetic Environment – Guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with syn-
thetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
Mains power quality should be that of a
typical commercial or hospital environ-
ment.
(2)