Honeywell DirectLine DL422 User Manual page 19

Sensor module for meredian ii orp electrodes
Table of Contents

Advertisement

Step
Operation
3
100 % Range Value
Selection
Edit 100 % Range Value
Selection
Select 100 % ORP
Value
Save the New 100 %
Range Value
4
0 % Calibration
Adjust 0 % Calibration
Save 0 % Calibration
5
100 % Calibration
Adjust 100 %
Calibration
Save 100 % Calibration
6
Return to Online Display
1/03
Press
rnGH (for 1 second) then,
(value of current 100 % Range Value Selection)
Momentarily
MODE
Flashing Display – You are now in EDIT mode
Hold
(value of current 100 % selection)
Selected 100 % ORP Value in 10 mV increments
Momentarily
Range: –1600 to +1600 mV (default 1600 mV)
MODE
(New Value)
Momentarily
AdJL
Momentarily
MODE
AdJL (flashes) – You are now in EDIT mode
Hold
Range: 3.80 to 4.40 mA typically (default 4.00 mA)
+AdJL (increments value)
−AdJL (decrements value)
Momentarily
MODE
AdJL
Momentarily
AdJH
Momentarily
MODE
AdJH (flashes) – You are now in EDIT mode
Hold
Range: 19.60 to 20.40 mA typically
(default 20.00 mA)
+AdJH (increments value)
−AdJH (decrements value)
Momentarily
MODE
AdJH
Momentarily
MODE
Returns to Online Display
Momentarily
DirectLine® DL422 Sensor Module User Manual
Configuration
Display
13