Understanding Test Results and Messages
The
GE006 Blood Glucose
Meter displays results between 10 and 600 mg/dL.
If the test result is below 10 mg/dL , "
" will appear on the screen. Please
repeat your test with by a new test strip.
If you still get a "
" result, contact your healthcare professional.
If the test result is above 600 mg/dL, "
" will appear on the screen. Please
repeat your test again with a new test strip.
If you still get a "
" result, contact your healthcare professional.
36
What is a Quality Control Test ?
To ensure proper monitoring function, it is necessary to regularly perform a quality control test.
Use one of the
Rightest™
Control Solution GC550 when testing your GE006 Blood Glucose Monitoring
System in the Control Solution Mode. If the test result is within the Control Solution Range printed on the
strip vial label, the GE006 Blood Glucose Monitoring System passes the Quality Control Test and your
GE006 Blood Glucose Monitoring System is working properly.
Control Solution Range:
Example of Control Solution Range printed on
your test strip vial label.
NOTE
- Level 2 of
Rightest™ Control Solution GC550
purchase
Level 4
or another
Level 2
control solutions,
Service at 886 0800 241 688.

About Quality Control Testing

Control Solution Range mg/dL
Level 2:
107-145
Level 4:
315-426
is provided in your starter kit. If you want to
please contact Bionime Customer
37