5.9 Manual Offset?

The Manual Offset? option enables you to apply a manually-entered offset (in percent or PPM) to the gas calculations.
FACTORY SETUP MENU
Manual Offset?
FACTORY SETUP MENU
% Offset: [X.XX]:
FACTORY SETUP MENU
Manual Offset?

5.10Change AutoCal Limits?

The Change AutoCal Limits? option enables you to adjust the error handling limits for auto calibration. (To program
basic auto cal parameters, refer to Auto Cal Parameters on page 42.) The TMO2D has default values for error
handling:
Total Calibration Drift Limit: 40% of input range (6.40 mA)
Drift per Calibration Limit:
Verification Mode Drift Limit: 10% of input range (1.6 mA)
FACTORY SETUP MENU
Change AutoCal Limits?
AUTOCAL DRIFT LIMITS
Edit Tot Drift Limit?
Set Total Cal Drift:
Drift mA [X.XX]:
AUTOCAL DRIFT LIMITS
Edit Tot Drift Limit?
AUTOCAL DRIFT LIMITS
Edit per cal Limit?
Set Max Drift Per Cal:
Drift mA [X.XX]:
AUTOCAL DRIFT LIMITS
Edit per cal Limit?
TMO2D Digital Display User's Manual
Press [YES] to apply a manual offset to the gas calculations.
Use the numeric keys to enter the desired value (in % or ppm). Press [YES] to enter
the value, and [YES] again to confirm the entry.
Press [NO] to proceed to the Change AutoCal Limits? option.
25% of input range (4.00 mA)
Press [YES] to change the error handling limits for auto calibration.
Press [YES] to edit the total calibration drift limit.
Use the numeric keys to enter the desired limit. Press [YES] to enter the value, and
[YES] again to confirm the entry.
Press [NO] to proceed to the drift per calibration limit.
Press [YES] to edit the drift per calibration limit.
Use the numeric keys to enter the desired limit. Press [YES] to enter the value, and
[YES] again to confirm the entry.
Press [NO] to proceed to the verification mode drift limit.
Chapter 5. Advanced Programming
59