Compressor Manufacturer
Philips Respironics
1. Two piece Valve Cover
Assembly
2. External thermal cutoff
switch
3.
New inlet tube
assembly design
Inlet Tube Assembly Removal
1. Remove the back cabinet of the EverFlo.
2. Inspect the EverFlo compressor for visible wear, deterioration, broken components, and service labels.
3. Make note of the manufacturer of the compressor by using the Compressor Identification table in the
Detailed Description section of this field communication.
a. If the compressor is not manufactured by Philips Respironics, this procedure is not applicable.
b. If the compressor is manufactured by Philips Respironics, and the inlet tube assembly requires
replacement, continue with this replacement procedure.
4. Remove the Philips Respironics manufactured compressor from the EverFlo device.
Page 74
EverFlo Compressor Identification Table
Compressor Photo
LO TM
E
F
S
& T
VER
ERVICE
1
2
I
ECHNICAL
NFORMATION
3
1039055, VER. 06