Hygromaster
Protimeter Thermo-Hygrometer
Instruction Manual

Advertisement

loading

Summary of Contents for GE Hygromaster

  • Page 1 Sensing ® Hygromaster Protimeter Thermo-Hygrometer Instruction Manual...
  • Page 2: Table Of Contents

    Logging Functions - Reading the Data ..........19 GE Protimeter Hygromaster...
  • Page 3: Introduction

    Introduction The GE Protimeter Hygromaster is a handheld hygrometer with data logging capability. Spot measurements of relative humidity, dew point, mixing ratio and ambient temperature are shown on a split display at the touch of a button. Surface temperature can also be measured using an optional surface temperature sensor.
  • Page 4: Hygrostick® And Quikstick Humidity Probe Options

    ® Hygrostick and Quikstick Humidity Probe Options The Hygromaster is used with a detachable humidity probe to measure relative humidity and ambient temperature. Two humidity probes are available, the Hygrostick and the Quikstick. The Hygrostick is optimized for measuring humidity in the 30-100 %rh range and may be used with Protimeter humidity sleeves for measuring the equilibrium relative humidity of solid walls and floors.
  • Page 5: Function Buttons

    Function Buttons The Hygromaster has four buttons (see Table 2 below) that are used individually or in pairs to take measurements, log readings and set-up the instrument. When pressed individually they perform the following standard functions: Table 2: Function Buttons...
  • Page 6: Standard Functions - Humidity And Temperature Measurements

    Standard Functions – Humidity and Temperature Measurements Measure humidity and temperature as follows: Connect the Hygrostick or Quikstick probe to the Hygromaster instrument either directly into the socket on the top of the instrument or indirectly with the optional extension lead.
  • Page 7: Function Tables

    Should successive readings on both halves of the display not change, the upper display Note: decimal is flashed to show that new readings are still being taken. Function Tables See Table 3 on the next page for Spot Measurements. GE Protimeter Hygromaster Page 7...
  • Page 8 Stores measurement in Press and hold for 3 seconds to store the instrument memory all measurements. Momentarily displays the log record number and stored data. GE Protimeter Hygromaster Page 8...
  • Page 9 - see Instrument Set Up (Table 7). Short cut to Switches surface temperature socket ON, OFF (RS232 mode) or Surface Auto mode. Temperature set (RS 232 mode) & up modes GE Protimeter Hygromaster Page 9...
  • Page 10 Returns instrument to normal operation without saving new settings. & Zeroes current parameter Resets the active parameter to minimum settings value (none) Timeout If no button is pressed for 5 seconds setup mode exits without saving changes. GE Protimeter Hygromaster Page 10...
  • Page 11 Logging Logging interval is set to 0 (OFF) 24 hr 15 minutes (example) WB First Delay before logging is set to 0 (no 24 hr none (take a log immediately delay) when settings accepted) GE Protimeter Hygromaster Page 11...
  • Page 12 ‘0’, logging will be active and the icon set. Pressing , or not pressing any key at all for five seconds at any point in the process will exit immediately without making any changes. GE Protimeter Hygromaster Page 12...
  • Page 13 Reset all instrument parameters to 0 (no) 0 (no) 1 (yes) factory Default Set automatic ShutDown Time. 2 mins 9 mins 0 = stay on until turned-off manually °C and ppmw, or °F and Select °C °C °F GE Protimeter Hygromaster Page 13...
  • Page 14 , then the instrument returns to normal operating mode without changing the set-up parameters. is pressed after ‘ ’, the new settings are accepted; except, the settings are retained when the instrument is switched OFF. GE Protimeter Hygromaster Page 14...
  • Page 15 Surface Temperature port Note: without using the instrument setup menu. Each time the key-pair is pressed, the lower display switches between ON and OFF Set side-port surface temperature ON or OFF GE Protimeter Hygromaster Page 15...
  • Page 16: Logging Functions - Pc Setup

    Confirm that the instrument serial number has been read correctly (not shown as “-”). This would indicate that the communications have been established. If not, check settings under the “Communication” tab. Click on “Set Time” to set the instrument clock to the PC clock time. GE Protimeter Hygromaster Page 16...
  • Page 17 Remove the RS232 serial cable from the instrument port and place it in the environment that is to be monitored. Ensure that a Hygrostick or Quikstick sensor is plugged into the top of the instrument. until the display shows OFF. Press and hold GE Protimeter Hygromaster Page 17...
  • Page 18: Logging Functions - Manual Logging (Recording)

    In this case, note that the side port must be previously set in “surface temperature” mode if this input is to be used. Again there is no requirement for logging to have been enabled first, although it may be useful to set the job number. GE Protimeter Hygromaster Page 18...
  • Page 19: Logging Functions - Reading The Data

    When the download completes, click on OK then save as prompted. Use the graphing tool for a quick inspection (select a subset of the records first, if required) or open the saved data file in a spreadsheet for further manipulation. GE Protimeter Hygromaster Page 19...
  • Page 20 1100 Technology Park Drive Billerica, MA 01821-4111 Web: www.gesensing.com/protimeterproducts Ireland Sensing House Shannon Free Zone East, Shannon, County Clare Web: www.gesensing.com/protimeterproducts ® ® Protimeter Hygromaster and Hygrostick are registered trademarks of GE Sensing. INS7700, Rev. B August 2007...