  • Page 1 User Instruction Manual Oxford Standaid 140 ® To avoid injury, read user manual prior to use.
  • Page 2: Table Of Contents

    Oxford ® English Standaid 140 Contents Introduction: About Your Lift ....................3 Assembly and Commissioning Instructions ............... 4 Safety Precautions ......................6 Control Box & Handset ...................... 8 Operating Instructions ....................... 9 Charging Instructions....................... 11 Maintenance Schedule & Daily Check List ..............13 Maintenance, Inspection and Test ..................
  • Page 3: Introduction: About Your Lift

    Recycling of Packaging: In an effort to reduce levels of waste and environmental impact, Joerns Healthcare operate a free of charge recycling service in the UK. This is specifically for the foam inserts that accompany the lift packaging - inside of the carton.
  • Page 4: Assembly And Commissioning Instructions

    Oxford ® English Standaid 140 2. Assembly and Commissioning Instructions Place the carton in a clear working area and open carefully. The carton contains: • Wheeled chassis • Mast/boom assembly • Wallet containing documents • Battery charger • Leg adjusting handle •...
  • Page 5 Oxford ® English Standaid 140 remove the screw from the leg adjusting lever located at the rear of the chassis. fit the leg adjusting handle through the gate on the mast and push the open end over the leg adjusting lever. The handle is retained by a screw through the clearance hole in the handle into a threaded hole in the lever.
  • Page 6: Safety Precautions

    Oxford ® English Standaid 140 3. Safety Precautions Please read and follow the safety precautions listed below. The operation and use of Oxford patient lifts is simple and straightforward. Following these few basic safety precautions will make lifting operations easy and trouble free. WARNING •...
  • Page 7 • DO NOT bump the lift down steps, loaded or unloaded. • DO NOT attempt to negotiate a loaded lift on a slope which exceeds 1:12 (approximately 5 degrees). Joerns Healthcare recommend a second helper is present when moving a patient on a slope.
  • Page 8: Control Box & Handset

    Oxford ® English Standaid 140 4. Control Box & Handset Handset Detachable Fig. 4 Battery Pack emergency Stop Button emergency raise/Lower Battery Level Indicator Fig. 5...
  • Page 9: Operating Instructions

    Oxford ® English Standaid 140 5. Operating Instructions Leg adjustment: The legs on the Standaid are adjustable for width. The legs can be opened to enable access around armchairs or wheelchairs. for transferring and negotiating narrow doorways and passages the lift legs should be in the closed position. To achieve the adjustment, the leg adjuster handle, located at the rear of the mast, is moved out, away from the mast and then sideways from rIGHT (fully closed) to LefT (fully open).
  • Page 10 NOTE: for detailed fitting instructions, please refer to the user guide supplied with each sling. WARNING Joerns Healthcare recommends that slings be checked regularly and particularly before each use for signs of fraying or damage. DO NOT use slings that are worn or damaged.
  • Page 11: Charging Instructions

    The batteries are located in the power pack and are charged through two contacts on the base. When the power pack needs charging it is removed from the lift and fitted to a charging unit. Joerns Healthcare recommend an additional power pack is purchased, so that one pack can be on charge at all times.
  • Page 12 Oxford ® English Standaid 140 WARNING NEVER run the batteries completely flat. As soon as the audible warning sounds, complete the lifting operation in hand and place on charge. NEVER store the power pack for long periods without regular charging throughout the storage period.
  • Page 13: Maintenance Schedule & Daily Check List

    NOTE: The LOAD TeST and CerTIfICATION should only be carried out by qualified personnel or an authorised service dealer. Daily Check List: Joerns Healthcare strongly recommend the following checks are carried out on a daily basis and before using the lift. •...
  • Page 14: Maintenance, Inspection And Test

    Standaid 140 8. Maintenance, Inspection and Test Joerns Healthcare recommend a thorough inspection and test of the Oxford Standaid and lifting accessories, slings etc. is carried out every six months. The examination and test should be conducted according to the recommendations and procedures below. Joerns Healthcare recommend maintenance, inspection and certified testing is carried out by authorised service dealers only.
  • Page 15 Oxford ® English Standaid 140 • SLINGS: Check for wear and fraying. DO NOT use any sling that is damaged, or if the sling shows signs of wear. • LOAD TEST: The load test should be carried out in accordance with the manufacturer’s test procedures.
  • Page 16: Technical Specifications

    Oxford ® English Standaid 140 9. Technical Specifications Safe Working Load ......................140kgs Maximum Overall Length ................... 1220mm Minimum Overall Length .................... 1170mm Maximum Overall Height .................... 1625mm Minimum Overall Height ..................... 1138mm Spreader Bar Max. Height ..................1585mm Spreader Bar Min. Height ................... 1090mm Height At Maximum reach ..................
  • Page 17 Oxford ® English Standaid 140 Electrical Specifications: Batteries ............2 X 12 volt rechargeable sealed lead acid type Battery Capacity ..........2.9 Ampere hours Charger rated Input ........230Vac 50/60 Hz Charger rated Output ........27.4/29.0 VDC@0.8A Electric Shock Protection: Charger............Class II Lift ..............Internal Power Source Degree Of Shock Protection: Charger............Type B Lift ..............Type B...
  • Page 18 Oxford ® English Standaid 140 Key Symbols: The following symbols are used on the charger, control unit and battery: Type B equipment, as per eN 60601-1 Class 2 equipment The disposal of the charging and control unit should not be mixed with general household waste The disposal of batteries should not be mixed with general household waste...
  • Page 19: Warranty

    If you are in doubt as to where your product was purchased, Joerns Healthcare can trace the supplier if you quote the serial number of the lift. Contact details can be found on the rear cover of this manual.