In stru ctio ns fo r Use
Data Export Test Tool
MP2/X2, MP5/20/ 30/40 /50/60 /70/80 /90 & MX8 00

Advertisement

Table of Contents
loading

Summary of Contents for Philips MPX2

  • Page 1 In stru ctio ns fo r Use Data Export Test Tool MP2/X2, MP5/20/ 30/40 /50/60 /70/80 /90 & MX8 00...
  • Page 2 Part Number 453564254321 Printed in Germany 01/11 *453564254321*...
  • Page 3: Table Of Contents

    Table of Contents 1 About this Guide About the Data Export Test Tool Manufacturer's Information Trademark Information 2 DETT - Instructions for Use Installation of DETT Un-Installation of DETT Deletion of Log Files System Requirements for DETT How to start DETT The Menu System List Window System Detail Window...
  • Page 5: About This Guide

    These Instructions for Use are for technicians and biomedical engineers using the Data Export Test Tool (DETT) with the Philips IntelliVue MP2 / X2 / MP5 / MP20 / MP30 / MP40 / MP50 / MP60 / MP70 / MP80 / MP90 /MX800 (M8102A / M3002A / M8105A / M8001A / M8002A / M8003A / M8004A / M8005A / M8007A / M8008A / M8010A / 865240) patient monitors, hereafter referred to as the patient monitor.
  • Page 6: Manufacturer's Information

    Philips shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance or use of this material.
  • Page 7: Dett - Instructions For Use

    DETT - Instructions for Use Installation of DETT Unzip the downloaded version to the desired installation directory. The following files are unzipped to the installation directory: \DETT BinaryStreamer.dll configuration.xml DataExportTestTool.chm DETT.exe MdilDataTypes.dll Tnlscat.x PluginInterfaces.dll \DETT\Plugins ConnectionStatePlugin.dll DataStorePlugin.dll DiscoverPlugin.dll DiscoverSerialPlugin.dll Un-Installation of DETT Delete the DETT installation directory including DETT.exe.
  • Page 8: Deletion Of Log Files

    2 DETT - Instructions for Use Deletion of Log Files DETT generates log files when connected to an IntelliVue monitor. Logfiles are marked as "*.txt" - Files with DevIFTest as prefix, e.g. DevIfTest_20100222_103626_COM1_115200.txt To delete the log files, manually delete the files that are either stored in the default program directory or the selected log files path (refer to "File ->...
  • Page 9: How To Start Dett

    2 DETT - Instructions for Use How to start DETT Double click on the "DETT.exe" file stored in the DETT installation directory to start the application. Windows system admin rights are required. After starting the application, the main window is displayed. No system data is displayed. The application window consists of these main areas: 1 - Menu 2 - System List Window...
  • Page 10: The Menu

    2 DETT - Instructions for Use The Menu File -> Configuration In the "File" pull down menu select "Configuration" to import a new NLS text catalog file, change the default path for the log files or change the baud rate of the COM interface. When selecting "Change path"...
  • Page 11 Documentation for the patient monitors. The NLS text catalog is included on the IntelliVue Support Tool DVD or can be downloaded from Philips InCenter. When clicking the "Import" button, a windows file browser menu opens and the Tnslcat.x file to be...
  • Page 12 2 DETT - Instructions for Use File -> Exit From the "File" file pull down menu select "Exit" to exit the DETT. This will terminate the program, including all open threads.
  • Page 13 2 DETT - Instructions for Use Start Ensure that you have connected the IntelliVue monitor to the PC, either via LAN (monitor must have IP address either via BootP server or manual configuration) or via RS-232. When using the LAN interface of the PC without a BootP Server in the LAN network (e.g. direct LAN Connection), the PC must use a valid IP address within the same subnet as the IV.
  • Page 14 2 DETT - Instructions for Use Stop To stop the data export communication select "Stop" from the Menu. The connection to all monitors is terminated. Logging is stopped as well. ? -> Help/About In the "?" pull down menu, "Help" and "About" can be accessed. Help Select "Help"...
  • Page 15 2 DETT - Instructions for Use Status Strip The Status Strip shows the current status of the DETT. The following status messages are displayed: Ready DETT is launched and ready to be started Running DETT has been started and is communicating Stopped DETT has been stopped...
  • Page 16: System List Window

    2 DETT - Instructions for Use System List Window After starting the DETT, the data export communication of the available patient monitor systems is displayed in the system list window. The system window provides the following information / selections. Debug This check box enables or disables the debugging feature of the data export protocol.
  • Page 17 2 DETT - Instructions for Use 1 – Connect indication request for the LAN interface. The Connect Indication message is only available on the LAN interface. As soon as the IntelliVue monitor has received a valid IP address from the BOOTP server in the network, the monitor sends out the Connect Indication message of its LAN interface.
  • Page 18: System Detail Window

    2 DETT - Instructions for Use System Detail Window The System Detail Window consists of several parts. Per detected patient monitor, a tab is added to the System Detail Window. The respective tab can be identified by the monitor label, the COM label or an empty tab.
  • Page 19 2 DETT - Instructions for Use Selectable Buttons within a Tab Copy to Clipboard When selecting copy to clipboard, the system details (system information and numeric data) will be copied to a clipboard. This allows export of data into e.g. MS Excel. The following details are copied and transferred to a spreadsheet.
  • Page 20 2 DETT - Instructions for Use Patient Data When selecting patient data, a new window will be opened displaying the patient data. Those details are not copied to the clipboard. For details on the ID attributes, refer to latest revision of the Data Export Interface Programming Guide.
  • Page 21 2 DETT - Instructions for Use MDS Data When selecting MDS data a new window will be opened displaying the connected patient monitor system details. For details on the ID attributes, refer to latest revision of the Data Export Interface Programming Guide.
  • Page 22 2 DETT - Instructions for Use Wave Priority List When selecting the Wave priority list, a new window will be opened displaying the current available Wave parameters and data transmission. For details on the ID attributes, refer to latest revision of the Data Export Interface Programming Guide.
  • Page 23: Debug Information Window

    2 DETT - Instructions for Use NOTE IntelliVue Rev. G.0 or higher allows the configuration of either a second MIB/RS232 port or the combination LAN interface and MIB/RS232 port for Data Export. The DtOut2 driver is used to connect a second port to Data Export. Only one connection is able to request wave data at a time, the other connection responds with a notification that wave polling is not possible.
  • Page 24: Log Files

    2 DETT - Instructions for Use Log Files The communication between the DETT and the Patient Monitor is logged in the selected log path. A log file is created each time the data export communication is started (refer to chapter "Start" for more details on how to start).
  • Page 25: Dett Overview Example

    2 DETT - Instructions for Use DETT overview example The screen shot below shows the DETT with several connected patient monitors. In this example one device is selected for debugging. The parameter values are displayed in the System Information section and the data stream details are visible in the Debug Information section of the tool window.
  • Page 26: How To Connect A Pc With Dett To An Intellivue

    2 DETT - Instructions for Use How to connect a PC with DETT to an IntelliVue The following chapter describes how to connect a PC using DETT to an IV or IV Network, either via LAN, LAN / RS-232 converter, RS-232 direct connection or USB/RS-232 adapter. WARNING All external devices in the patient vicinity must comply with IEC 60601-1:1988/A1:1991A2:1995 or EN 60601-1:1990/A1:1993/A2:1995.
  • Page 27: Index

    Index ? -> Help/About Manufacturer's Information MDS Data About About the Data Export Test Tool About this Guide Name COM Speed COM/MAC Patient Data Connecting an IntelliVue Monitor LAN Interface by using a LAN/RS232 converter Selectable Buttons within a Tab Connecting an Intellivue Monitor MIB/ RS232 Interface by using a USB / RS- Start...