Alarm Temperature Setting - Panasonic MDF-U4186S Series Operating Instructions Manual

Ultra-low temperature freezer
Table of Contents

Advertisement

ALARM TEMPERATURE SETTING

This unit is provided with the high temperature alarm. The setting of high temperature alarm is 10
o
15
C (The setting value is 2 kinds.) higher than the setting of chamber temperature. The procedure in
table 2 shows the sequence to set the high temperature alarm at 15
temperature. Set the figure to 000 at step 5 when setting the high temperature alarm at 10
than the setting of chamber temperature.
Note:
The alarm temperature is set at the factory 10
Table 2. High temperature alarm setting (15
Description of operation
1
Press the numerical value shift key
2
for about 5 seconds.
Set the first digit to 1 with the
3
numerical value shift key.
4 Press enter key.
Set the first digit to 1 with the
5
numerical value shift key.
6 Press enter key.
Note:
The setting is not memorized if the mode change key (PV/SV) is pressed during alarm
temperature setting. The chamber temperature is displayed for 3 seconds when the numerical value
shift is pressed 5 seconds continuously.
o
C higher than the setting of chamber temperature.
o
C higher than the setting of chamber temperature)
Key operated
-----
The current chamber temperature is
displayed.
The first digit blinks.
The first digit blinks.
ENT
The first digit blinks.
When pressed, the figure of settable
digit changes.
Alarm temperature is memorized and
ENT
the current chamber temperature
is displayed.
16
o
C higher than the setting of chamber
Indication after operation
o
C or
o
C higher