Scale Calibration Procedure (Lbs. & Kg) With Remote Display - Honeywell MS2320 Installation And User Manual

Stratosh scanner / avery scale
Hide thumbs Also See for MS2320:
Table of Contents

Advertisement

S
O
CALE
PERATION
C
ALIBRATION
Scale Calibration Procedure (lbs. or kg) with Remote Display*
1.
Temporarily remove the platter and place it in a safe location.
It is the responsibility of the owner of the scale to confirm compliance with the relevant Weights and Measures
statutes and regulations applicable in your area by checking with the appropriate government agency before
placing a newly calibrated unit into service or removing any official seals.
2.
If this is a currently installed scanner/scale* in need of calibration, cut and remove the calibration switch cover seal. If this
is a new installation, cut and remove the factory-applied adhesive seal.
Follow all Electo-Static Discharge (ESD) procedures when exposing internal scanner/scale components.
3.
Remove the M3 screw securing the calibration switch/button cover.
Place the cover and screw in a safe location. Verify that the scale calibration switch is in the Run position.
4.
Enter full service access mode. Power down the unit if necessary. Press and hold down the calibration push button
then power up the scanner/scale. Release the calibration push button.
See power source caution statement on page 8 of this manual.
* These procedures are for MS2320 StratosH Scanner/Avery Scale models only
38
Figure 41. Calibration Switch/Button Cover Removal
Figure 42. Entering Full Service Access Mode
.

Advertisement

Table of Contents
loading

Table of Contents