Switch operation
 
ESC 
● Changing the range type
The range type should be set before changing the span.
The following is an example to change the range type from Single range (SINGLE) to
Bidirectional automatic selection of multiple ranges (2F-2R).
Switch operation
 
STEP1 
 
 
 
STEP2 
 
 
 
STEP3 
 
 
 
STEP4 
 
 
 
Display example
C:RANGE 
Pushing
1 R TYPE 
menu.
2 R1  3 R2 
4 R3  5 R4 
6 R HYS 
7 EXIT 
 
ESC
 
Display example
C:RANGE 
Select "R TYPE" from the setting item
1 R TYPE 
selection menu.
2 R1  3 R2 
 
4 R3  5 R4 
 
6 R HYS 
7 EXIT 
 
ESC
 
R TYPE 
The currently set range type (SINGLE in
 
this example) appears.
SINGLE 
 
Then push
 
 
* Pushing
 
ESC 
 
 
R TYPE 
The switches at the bottom change.
 
(
4F-0R 
At the same, the cursor appears.
 
(The item indicated by the cursor is
        
highlighted.)
 
Then push
R TYPE 
You can continue to change the setting item.
 
Push
2F-2R 
items.
 
        
 
 
R TYPE 
Perform this operation to change the setting
 
to 2F-2R.
2F-2R 
 
When the item is changed to the desired
 
item, push
temporarily.
− 69 −
Description
returns you to the setting
ESC
Description
.
returns you to the setting
ESC
menu.
are shown.)
to change the selection
: Selected item is scrolled up.
: Selected item is scrolled down.
to set the item