Centrifuge Module Interface Operating Instructions
12. Using Hettich lubricating grease, lubricate the cylinder wall of the rotor
pins.
Figure 12:
1.
2.
Cleaning the Hettich Unit Chamber and Seals
Cleaning Cycle
Materials Needed
To clean the Centrifuge chamber and seals, perform the following steps:
1. Inspect the outer and inner seals.
If a seal is damaged, call your local technical support provider for
replacement seals.
092D0001-02 Rev. A
Rotor and Pins
Rotor
Cylinder wall of the rotor pins
As needed
Cleaning cloth
10% bleach solution
Hettich lubricating grease
Isopropyl alcohol
Maintenance