GE AKTA pilot 600 User Manual page 90

Table of Contents

Advertisement

5 Maintenance
Precautions
90
WARNING
Electrical shock hazard. All repairs should be done by service
personnel authorized by GE. Do not open any covers or replace
parts unless specifically stated in the user documentation.
WARNING
Use only approved parts. Only spare parts and accessories that
are approved or supplied by GE may be used for maintaining or
servicing the product.
WARNING
For continued protection against injury risks due to fluid jets, burst
pipes or potentially explosive atmosphere, the user must test the
piping system for leakage at maximum operating pressure.
Always perform a leakage test after assembly or maintenance.
Always perform a leakage test before operation or CIP.
WARNING
Hazardous chemicals during run. When using hazardous chemi-
cals, flush the entire system tubing with distilled water, before ser-
vice and maintenance.
WARNING
Decontaminate before maintenance. To avoid personnel being
exposed to potentially hazardous substances, make sure that the
column is properly decontaminated and sanitized before mainte-
nance or service.
ÄKTA pilot 600 User Manual 29274559 AA

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Akta pilot 600sAkta pilot 600r

Table of Contents