Adjustments / Programs
Release fluoroscopy. The X-Iris must close between 4 - 4.5 cm.
Stop fluoroscopy.
Select "5. CLOSED POSITION" on the service PC and click on "Execute".
Close the status window by clicking on the OK button.
For the query "Do you want to reset the unit," click "No".
Select "8. Init Off & Store Positions" on the service PC and click on "Execute".
Close the status window by clicking on the OK button.
For the query "Do you want to reset the unit," click "Yes".
Switch the SIREMOBIL Compact OFF and ON again.
Check the X-Iris again in survey format and zoom format. The collimator blades must still
be visible on the edge of the image. If necessary, repeat the adjustment.
Stop fluoroscopy.
Return to the main menu on the service PC with the ESC key.
Select the Data, Backup... menu.
Select Parameter in the backup window.
Click on "Backup".
At "Remarks for Parameter" enter "X-Iris set".
Click on "OK". The new position values for the collimator will be stored on the disk.
Return to the main menu with the ESC key.
Siemens AG
Medical Engineering
Register 5
RXR2-130.061.02
Rev. 03
Page 11 of 22
10.03
CS PS 24
5 - 11
SIREMOBIL Compact
System Manual