Prerequisites
I.I. cover
The cover for the image intensifier is secured by one screw.
Insert the rubber lip in the groove prior to remounting the cover.
When installing or deinstalling, handle the protective conductors with care.
I.I. mini power supply
The control panel of the image intensifier mini voltage supply is accessible after
removing the I.I. cover.
The high voltage panel of the I.I. mini power supply is built in to the C-arm (see Fig. 1).
The cover (on the outside of the C-arm near the I.I.) is secured with two screws and can
be removed from the rubber buffer of the I.I. After removing the cover, the high voltage
panel is accessible.

SIREPHOS cover

NOTE
Loosen the cover screws of the SIREPHOS cover.
Use a sharp knife to cut open the sealing compound all around the cover.
Lift off the SIREPHOS cover.
Remove the sealing compound residue from the cover and from the SIREPHOS.
On completing service, place the SIREPHOS cover on the SIREPHOS and seal it with
sealing compound all around.
Use a paper towel to wipe off excess sealing compound.
Refasten the SIREPHOS cover with the cover screws.
Siemens AG
Medical Engineering
The SIREPHOS cover has been sealed all around against contam-
ination by dirt using a sealing compound (Part No. 20 49 716).
Register 5
RXR2-130.061.02
Rev. 03
Page 7 of 10