By entering into this program the last value displayed before turning OFF the Unit will
be displayed on turn it On the next time .

7.17.6 Program 6

To Display the maimum KV & mAS possible with the Unit
100 indicates the maximum KV available & 200 indicates the maximum mAS
available on the Unit
*In case the display values do not vary in standby by using the key PCB then
Enter this program and vary the KV & mAs displayed using the Key PCB D936
Then switch the the Unit O FF and enter
normally .

7.17.7 Program 7

Adjusting REG frequency.
Please refer chapter 7.8 (Checking & setting the maximum main inverter frequency)