3-4-3. Replacing the Filter Assembly

Removal
1. Turn OFF the AC power supplied to the FVS-1000.
n
Be sure to turn off the AC power supplied to the FVS-1000
first before proceeding with the work due to the risk of
electrical hazards.
2. Remove the gate base (35) assembly.
(Refer to section 3-2-1.)
3. Loosen the four screws, and pull out the lamp house
assembly. (Refer to section 3-4-1.)
4. Remove the LC-35 board cover.
5. Disconnect all connectors connected, and remove the
LC-35 board.
LC-35 board
LC-35 board cover
FVS-1000 MMP1
PSW3 x 6
PSW3 x 6
B3 x 6
3-4. Replacement of Periodic Replacement Part
6. Remove the six screws, and remove the lamp house
front panel.
7. Remove the four screws, and remove the filter assem-
bly.
Filter assembly
Lamp house panel
assembly
B3 x 12
B3 x 12
B3 x 12
Installation
8. Perform the reverse of steps 1 to 7 of the removal
procedure.
n
Refer to the following diagram for the connection of
connectors to the LC-35 board.
LC-35 board
CN101
CN21
CN20
CN1
CN19
CN2
CN4
CN18
CN17
CN16
CN5
CN7
CN503
CN6
B4 x 14
B4 x 14
CN201
CN15
CN3
CN14
CN11
CN9
CN12
CN8
CN10
CN13
3-17