Mechanical systems
Friction lining
Fig. 4 For upper arm stability
!
Adjustment screw
Fig. 6
Counterbalance
•
Remove the plastic plugs covering the friction linings in the column.
•
Loosen the friction linings through the holes in the column (Fig. 4) and the lining in the
lower joint (Fig. 5).
•
Check that the arm system stops in any position.
If does not stop, but moves lower or higher:
•
Adjust the arm system with the adjustment screw (Fig. 6) until it balances in all positions.
If the arm moves lower, the screw must be tightened; if it moves up, the screw must be
loosened.
Siemens AG
Medical Solutions
Friction lining
Fig. 5 For lower arm stability
SPR8-220.061.03
Page 3 of 28
Rev. 03
09.03
CS PS 24
10 - 3
10
MOBILETT Plus HP