Mechanical systems
Friction lining
Fig. 4 For upper arm stability
!
Adjustment screw
Fig. 6

Counterbalance

Remove the plastic plugs covering the friction linings in the column.
Loosen the friction linings through the holes in the column (Fig. 4) and the lining in the
lower joint (Fig. 5).
Check that the arm system stops in any position.
If does not stop, but moves lower or higher:
Adjust the arm system with the adjustment screw (Fig. 6) until it balances in all positions.
If the arm moves lower, the screw must be tightened; if it moves up, the screw must be
loosened.
Siemens AG
Medical Solutions
Friction lining
Fig. 5 For lower arm stability
SPR8-220.061.03