Siemens MOBILETT Plus HP Service Instructions Manual page 74

Hide thumbs Also See for MOBILETT Plus HP:
Table of Contents

Advertisement

8 - 8
2
1
Fig. 10
Disconnect the wires (pos. 1 / Fig. 10) and mark if necessary.
Connection to Connection from Color
S11-2
S11-3
S11-10
S11-12
S11-12
S11-13
S11-14
Replace switch element (pos. 2 / Fig. 10).
Reinstallation is performed in the reverse order.
NOTICE
MOBILETT Plus HP
MOB00624
BATT+
RD
K202:3
BU
K52:3
BK
K104:1
RO
K202:2
YE
Fuse U10
BK
K201:1
BK
During reinstallation pay special attention to the shelf of the
power transmission and the rail of the cable winder assembly,
which must be under the shelf!
We recommend that you pull out the power transmission unit to
make the reinstallation easier.
SPR8-220.061.03
Rev. 03
09.03
Miscellaneous
Connection to Connection from Color
S11-15
K201:3
S11-16
Fuse U9
S11-19
S11-21
K104:4
S11-21
S11-23
K301-2
S11-24
K301:1
Page 8 of 8