Testing Xgm868I Software And Hardware; Uploading Signal Array Data - GE DigitalFlo XGM868i Service Manual

Panametrics general-purpose gas flow transmitter (1 & 2 channel)
Table of Contents

Advertisement

Appendix B. Calibration and Testing with PanaView
B.5 Testing XGM868i Software and Hardware
In addition to calibration procedures for input and output cards, the Calibration/Test menu offers tests for XGM868i
hardware and procedures for uploading signal array data and XGM memory contents into PC files for examination. To
test XGM868i hardware and software,
1. Open the Meter Browser window and highlight the XGM868i.
2. Expand the Edit Functions option. The window now appears similar to that shown in Figure 9 on page 42.
3. Double-click on the Calibration/Test menu. The window appears similar to Figure 10 on page 42.
Proceed to the appropriate section to upload data and test hardware.
B.5.1 Uploading Signal Array Data
By uploading XGM868i signal array data for examination, a user can check if the transducer signal is the proper shape,
and change setup parameters if necessary.
Consult GE before using this data to change any parameters.
IMPORTANT:
1. In the Calibration/Test menu, double-click on the Signal Array Upload option.
2. Double-click on the desired Channel.
3. The next prompt asks for the specific Array to Load. Five choices are available:
Raw UP (Upstream) Signal
Raw DOWN (Downstream) Signal
CFUNC UP (Correlation Function Upstream)
CFUNC DOWN (Correlation Function Downstream)
CCFUNC (Cross-Correlation Function)
Double-click on the desired array.