TRIGger Subsystem
CONDition
When LT (less than) is selected, the oscilloscope will trigger on the first
transition that causes the pattern specification to be false, after the pattern
has been true for the number of times specified by the trigger event count
(DELAY command). The first event in the sequence will occur when the
specified pattern is true for a time less than that indicated by the trigger
specification. All other pattern true occurrences in the event count are
independent of the pattern duration time.
real number between 20 ns and 160 ms
<time>
Example
OUTPUT XXX;":TRIG:COND ENT"
The oscilloscope cannot be programmed for a pattern duration (GT, LT, or
RANge) trigger if it is being armed by another module via an IMB (Intermodule
Bus) measurement.
Query
:TRIGger:CONDition?
The CONDition query returns the present condition.
Returned Format
[:TRIGger CONDition]
{ENTer|EXIT|GT,<time>|LT,<time>|RANGe,<time>,<time>}<NL>
Example
OUTPUT XXX;":TRIG:COND?"
9–6